# Analyze a run you already have

> Import a run file your cycler already saved, then check its curves, efficiencies, reference genes and data health. All you need is the file.

Source: https://clarida.bio/help/import-your-own-run
Updated: 2026-11-03

[Help](https://clarida.bio/help)

Import a run file your cycler already saved, then check its curves, efficiencies, reference genes and data health. All you need is the file.

By [Ward Hellemans](https://clarida.bio/help/import-your-own-run#about-the-authors)

November 3, 2026 3 min read

The run file may be on the instrument PC, a network drive or a USB stick. Import it as the instrument saved it: no export, no reformatting.

| From | File | What comes in |
| --- | --- | --- |
| **Bio-Rad** CFX instruments | `.pcrd` | Plate layout, amplification and melt curves, instrument Cq |
| **Roche** LightCycler 480 | `.ixo .xml` | Plate layout with samples, amplification curves; targets, melt curves and the instrument’s Cq when the file holds them, otherwise Clarida calls Cq |
| **Thermo Fisher** QuantStudio and ViiA 7 (not the older 7500 and 7900HT files) | `.eds` | Plate layout, amplification and melt curves, instrument Cq |
| **QIAGEN** Rotor-Gene | `.rex .qrex` | Tubes listed by number with their samples, amplification curves, melt curves when run; the file has no Cq, so Clarida calls it |
| **RDML** The open data standard for qPCR, written by many instruments and tools | `.rdml .rdm` | Whatever the file holds: annotation, curves, Cq |
| **Any instrument** A Cq table exported from its software | `.xlsx .csv .txt .tsv` | Cq values and whatever annotation the table carries; no curves |

Any other instrument: export its Cq table (`.xlsx`, `.csv`, `.txt` or `.tsv`) with at least a well and a Cq column; sample and target names come along when it has them. [Importing qPCR run files and Cq (Ct) tables](https://clarida.bio/methods/importing-instrument-data) lists the column names Clarida recognizes.

The screens below show a SYBR Green run from a Bio-Rad CFX96: Actin, IL1Beta and Tubulin, at 0Hr, 1Hr and 2Hr in triplicate, with a dilution series and a no-template control. Follow along with a run of your own.

![NRQ bar chart of IL1Beta at 0Hr, 1Hr and 2Hr: about 4.2, 1 and 0.25](https://clarida.bio/images/help/import-hero.png)

Figure 1: Where this guide ends: IL1Beta at 0Hr, 1Hr and 2Hr, normalized to Actin.

## How to

### 1. Start an experiment from your file

On your home page or the qPCR start page, choose **Start from data you already have**. The experiment skips Define, Design and Execute, since the file describes the plate, and opens on **Analyze → Import & annotate**.

![The qPCR start page's two Start new cards: Build from scratch, from Define to Report, and Start from data you already have, from Analyze to Report, with the link Analyze your data](https://clarida.bio/images/help/import-start.png)

Figure 2: Start from data you already have: the experiment begins at Analyze.

### 2. Drop the run file

Drag the file onto the drop zone, or click it to browse, then choose **Parse selected**.

![Import & annotate after the file was read: the steps Import, Link and Reconcile ticked, and one source, time-course.pcrd, Resolved, with its run](https://clarida.bio/images/help/import-done.png)

Figure 3: Import & annotate after the file was read: three steps ticked, one source.

**What to notice:** **Import**, **Link** and **Reconcile** are all ticked. They check a run against a plate laid out in Clarida; with none, the file's own annotation is used as it is, and the file becomes a run.

### 3. Check the curves and the imported Cq

Go to **Raw data → Amplification**, and use the arrows above the chart to step through Actin, IL1Beta and Tubulin. A Cq table has no curves, so this step is for run files.

![Raw data for Actin: amplification curves, fluorescence against cycle, the list of wells, and a table with each well's run, dye, sample, assay and imported Cq; the no-template control has none](https://clarida.bio/images/help/import-curves.png)

Figure 4: Raw data: Actin's amplification curves, and each well's imported Cq.

**What to notice:** the table keeps the Cq your instrument software called, under **Imported Cq**, and the no-template control (NTC) stays flat. Clarida can call Cq from the curves too, under **Cq method**; the imported values stay in use until you switch. A Rotor-Gene file holds no Cq, so there Clarida calls it.

### 4. Check the reference genes

Go to **Cq processing → Normalization**: the reference genes with their stability, and below them each sample's normalization factor.

![Normalization with one reference gene, Actin: mean Cq 22.44, detected in 100% of samples, a dash for M and a CV of 0.0%; the normalization factors of 0Hr, 1Hr and 2Hr all close to 1](https://clarida.bio/images/help/import-normalization.png)

Figure 5: Normalization on one reference gene, Actin.

**What to notice:** the CFX file marked Actin as a reference gene, so the run is normalized on it at once. With one reference gene, M shows a dash, and a CV of 0.0 says nothing about stability: there is no second gene to compare it with.

If the file doesn't name its reference genes (a Cq table and a Rotor-Gene file never do), Normalization lists your targets instead: tick them in the **Reference gene** column and choose **Confirm**.

### 5. Read each target's efficiency

Go to **Cq processing → Efficiency** and select a target to see its standard curve.

![Amplification efficiency: all 3 efficiencies within range; IL1Beta and Tubulin at E 1.94, Actin at 1.88, and the standard curve of Tubulin with E 1.94, R² 0.9997 and 18 points](https://clarida.bio/images/help/import-efficiency.png)

Figure 6: Amplification efficiency per target, and Tubulin's standard curve.

**What to notice:** each target has its own efficiency, from the run's dilution series: 1.94 for Tubulin and IL1Beta, 1.88 for Actin. Clarida turns Cq into relative quantities with each target's own value, instead of assuming perfect doubling.

### 6. Read the data health check

Go to **Exploration**. **Data health** at the top collects the checks the result rests on; the **NRQ bar chart** below shows one target at a time.

![Exploration: Data health with one orange line, 1 replicate set exceeds threshold, and three green; below it the NRQ bar chart of IL1Beta at 0Hr, 1Hr and 2Hr](https://clarida.bio/images/help/import-exploration.png)

Figure 7: Exploration: Data health with one orange line, and IL1Beta's result.

**What to notice:** one line is orange, **1 replicate set exceeds threshold**: an IL1Beta triplicate near Cq 30, its replicates 0.68 cycles apart, above the 0.5 set for this experiment. It asks you to look at that triplicate; it doesn't mean the analysis failed. Use the arrows next to **Target** to step to IL1Beta, the result in Fig. 1: about four times lower each hour.

## Try this

With your own run in view, change one thing and watch the analysis follow.

- **Look at the flagged triplicate.** Under **Cq processing → Replicates**, find the IL1Beta set above the threshold, then exclude its deviating well under **Cq processing → Exclusion**. Data health follows; include the well again to undo it.
- **Open the melt curves.** When the file holds them, **Raw data → Melt** shows them.
- **Add a second run.** Drop another file into **Import & annotate**, choose **Parse selected**, then **Promote all unlinked (1)**: the file becomes a second run of the same experiment.

## What you just did

- Brought a run into Clarida as your instrument saved it.
- Checked its curves next to the Cq your instrument software called.
- Saw which reference gene it is normalized on, and what one gene can't tell you.
- Read each target's efficiency and the data health of the result.

Which exports Clarida reads, and how a Cq table's columns are matched: [Importing qPCR run files and Cq (Ct) tables](https://clarida.bio/methods/importing-instrument-data).

## About the author

- Ward Hellemans

  Co-founder, Clarida

  Runs the product and the company side of Clarida.

## Stuck on your run file?

A founder answers within a working day. Say which instrument and software version wrote it, and where the import stopped.

[Start free](https://app.clarida.bio/login) · [Talk to a founder](https://clarida.bio/talk-to-us?topic=help)