# Plate homogeneity: check a cycler well by well

> Check a qPCR cycler for well-to-well variation: Cq values of a uniform 96- or 384-well plate in, pass or fail, outlier wells and a heatmap out. Free, no account.

Source: https://clarida.bio/tools/plate-homogeneity
Updated: 2026-11-02

[Free tools](https://clarida.bio/tools)

Load the Cq values of a 96- or 384-well plate with one reaction mix throughout, and see whether your instrument reads every position alike.

Free, no account

[Open the tool](https://app.clarida.bio/plate-homogeneity)

- **You start with**: Cq values of a uniform 96-well plate
- **You end with**: Pass or fail, outlier wells, heatmap

Video: The plate homogeneity tool on its demo plate: Results opens on an overall FAIL, with the variability test PASS and the outlier test FAIL; mean 25.59, median 25.52, SD 0.29, and 10 outlier wells. The heatmap shows row C in dark red as high outliers, with the row and column averages beside the plate. A click on well C9 opens it in the inspector: value 26.300, an outlier, against a row average of 26.206 and a column average of 25.535. In the settings, the maximum standard deviation goes from 1 to 0.25, and the variability test fails too. [Watch full size · 0:36](https://clarida.bio/videos/plate-homogeneity-hero.mp4)

Figure: Row C runs high: ten outlier wells fail the plate, and with a stricter SD limit its spread fails as well.

## What it tests, beside the published limits

Two tests on the whole plate; it passes when both do.

- **Variability test**: The standard deviation of all Cq values: at most 1 cycle by default. [The published protocol](https://clarida.bio/blog/functional-qpcr-instrument-validation) asks for 0.2 to 0.25 at a Cq around 25
- **Outlier test**: A well more than 0.5 cycles from the plate’s median is an outlier; the plate passes with at most 2
- **Your own limits**: Set each one in the settings, and the result updates
- **Left out**: Empty wells, and wells reading Undetermined or N/A

## How it works

01 In the tool: Plate homogeneity › Import data

### Load one plate’s Cq values

Upload one plate, 96-well (wells A1 to H12) or 384-well (A1 to P24), as an Excel, CSV or text file; a well written A01 or a1 reads as A1. Pasting from a spreadsheet works too. It needs the well and its Ct or Cq, knows the usual headers, and lets you map the columns when it does not. Cq values run from 0 to 99. No data at hand? Load the demo plate from the Help tab.

02 In the tool: Plate homogeneity › Results

### Read the verdict, then the plate

An overall pass or fail with each test’s outcome, the mean, median and standard deviation, and the outlier wells by position. The heatmap shows every well with its row and column averages beside it, so a warm edge or a drifting row stands out. Click a well for its value, its distance from the row, the column and the plate mean, and its Z-score.

03 In the tool: Plate homogeneity › Settings

### Hold the plate to your own SOP

Each limit can change, and the verdict follows.

- **Maximum standard deviation:** 1 cycle by default
- **Outlier distance:** more than 0.5 cycles from the median by default
- **Outliers allowed:** 2 by default

## What plate homogeneity doesn’t do

- **dPCR copy numbers**: Not yet: it reads a Cq column, with values up to 99
- **Keep the result with the run**: [Plate QC in Clarida](https://clarida.bio/qpcr/plate-qc) saves it with the experiment; here it stays on screen

## Frequently asked questions

### Is the plate homogeneity tool free?

Yes, with no account: open it, load your plate and read the result.

### How do you set up the test plate?

Pipette one reaction mix, for one target, into every well of the plate, with enough template for Cq values around 25, and run it as usual. [Functional qPCR instrument validation](https://clarida.bio/blog/functional-qpcr-instrument-validation) has the whole protocol.

### Can you use your own SOP’s limits?

Yes. The default standard deviation of 1 cycle is loose; set 0.25, or what your SOP says, and the verdict updates.

### Does it detect edge effects?

It shows them: the heatmap and the row and column averages make an edge that runs high or low easy to see. The pass or fail itself is for the whole plate, from its spread and its number of outlier wells.

### Does it count missing values?

Not in the pass or fail. Wells without a Cq stay out of the statistics and show as empty on the heatmap; the import says how many of the plate's wells have a value, so you can judge the rest.

## Cite this tool

Interactive on the page: Format: APA, Vancouver, BibTeX; Copy.

Clarida. (2026). Plate homogeneity (Version 2026.11.0) \[Web application\]. Retrieved November 18, 2026, from https://app.clarida.bio/plate-homogeneity

Updated November 2, 2026

## Run one uniform plate and read it here.

Load its Cq values; the verdict, the outlier wells and the heatmap follow.

[Open the tool](https://app.clarida.bio/plate-homogeneity) · [Plate QC in Clarida](https://clarida.bio/qpcr/plate-qc)