What it tests, beside the published limits
Two tests on the whole plate; it passes when both do.
| Variability test | The standard deviation of all Cq values: at most 1 cycle by default. The published protocol asks for 0.2 to 0.25 at a Cq around 25 |
|---|---|
| Outlier test | A well more than 0.5 cycles from the plate’s median is an outlier; the plate passes with at most 2 |
| Your own limits | Set each one in the settings, and the result updates |
| Left out | Empty wells, and wells reading Undetermined or N/A |
How it works
01In the tool: Plate homogeneity › Import data
Load one plate’s Cq values
Upload one plate, 96-well (wells A1 to H12) or 384-well (A1 to P24), as an Excel, CSV or text file; a well written A01 or a1 reads as A1. Pasting from a spreadsheet works too. It needs the well and its Ct or Cq, knows the usual headers, and lets you map the columns when it does not. Cq values run from 0 to 99. No data at hand? Load the demo plate from the Help tab.
02In the tool: Plate homogeneity › Results
Read the verdict, then the plate
An overall pass or fail with each test’s outcome, the mean, median and standard deviation, and the outlier wells by position. The heatmap shows every well with its row and column averages beside it, so a warm edge or a drifting row stands out. Click a well for its value, its distance from the row, the column and the plate mean, and its Z-score.
03In the tool: Plate homogeneity › Settings
Hold the plate to your own SOP
Each limit can change, and the verdict follows.
- Maximum standard deviation: 1 cycle by default
- Outlier distance: more than 0.5 cycles from the median by default
- Outliers allowed: 2 by default
What plate homogeneity doesn’t do
| dPCR copy numbers | Not yet: it reads a Cq column, with values up to 99 |
|---|---|
| Keep the result with the run | Plate QC in Clarida saves it with the experiment; here it stays on screen |
Frequently asked questions
Is the plate homogeneity tool free?
Yes, with no account: open it, load your plate and read the result.
How do you set up the test plate?
Pipette one reaction mix, for one target, into every well of the plate, with enough template for Cq values around 25, and run it as usual. Functional qPCR instrument validation has the whole protocol.
Can you use your own SOP’s limits?
Yes. The default standard deviation of 1 cycle is loose; set 0.25, or what your SOP says, and the verdict updates.
Does it detect edge effects?
It shows them: the heatmap and the row and column averages make an edge that runs high or low easy to see. The pass or fail itself is for the whole plate, from its spread and its number of outlier wells.
Does it count missing values?
Not in the pass or fail. Wells without a Cq stay out of the statistics and show as empty on the heatmap; the import says how many of the plate's wells have a value, so you can judge the rest.