What you get
A call table: one row per sample, one column per target, one call in each cell. Above it, three cards answer the questions in order: what was tested (samples × targets), whether the controls passed, and the result, such as “3 of 7 positive”. A Detected column counts the positives in each row. With controls configured, a QC dot on every row and column shows where a failing one points. Hover a cell to see its count, for example 2/3.
Save the analysis and it is kept as a frozen record, with the cut-off each target was called against. Add it to a PDF or HTML file with the Report builder, which carries the call grid and the counts.
How is a call made?
As a kit insert’s interpretation table states it, for every unknown sample and every target you test.
| Detected | Every included replicate has a Cq below the target’s cut-off |
|---|---|
| Not detected | No included replicate has |
| Inconclusive | The replicates disagree; Clarida does not average its way to a call |
| Not measured | Nothing is left to judge, so a gap never reads as a negative |
Where Clarida differs from most kit inserts: a failing control flags the result, and never changes the call.
| The internal control (IAC, often called IPC) fails in a sample | That sample’s row is flagged; its calls stay as computed, and there is no Invalid call |
|---|---|
| A positive or no-template control fails | The target it implicates is flagged, on its column |
| The IAC’s acceptable range | Its kit’s, else the experiment’s positive-control range; with neither, it must amplify |
How it works in Clarida
01In the app: Design › Assay Prep
A cut-off for each target, and which ones are controls
Each target carries its own detection cut-off, in Cq. Its type decides whether it becomes a column of the call table or a check on it. A gene of interest or a reference gene gets calls; a control is checked.
- Detection cutoff (Cq): per target; left empty, the experiment’s default applies
- Target type: gene of interest by default; the three control types (positive, negative, IAC) are checks
In full on the Design page
02In the app: Import › Import & annotate
Your run in, with every sample’s role
Import the instrument’s file or a Cq table, and each well is matched to its sample and assay. The type you give each sample matters here: only unknowns get a call.
- Checked, not called: positive, no-template, no-RT and other negative control samples
On the Import page: Import & annotate
03In the app: Analyze › Raw data
Curves to check before you trust a call
When the file brings curves, you see each well’s amplification and melt plot, and choose whose Cq the call uses. A late, noisy curve close to the cut-off is exactly the one worth a look.
- Cq source: automatic by default, the cycler’s when the file has it, otherwise Clarida’s, for the whole experiment
- Cq method: a curve fit per well (Clarida’s default), a baseline and threshold, the second-derivative maximum
In full on the Analyze page
04In the app: Analyze › Sample/Assay QC
Controls checked per sample and per target
The internal amplification control is checked in every unknown, so a sample where it failed is flagged. Positive control samples must amplify each target, and negatives must stay clean. The verdicts carry through to the call table.
- Negative controls: automatic by default, a Cq floor or a ΔCq margin
In full on the Analyze page
05In the app: Analyze › Detection
The call table, recomputed as you work
Detection computes as soon as you open it, and again whenever a Cq, an exclusion, a sample type or a cut-off changes. An excluded well leaves the call, which is then made on the replicates that remain. Save the analysis to freeze it.
- Detection Cq cut-off: the experiment’s default, empty unless you set it (any Cq counts); shared with Genotype calling
On the Report page: Report builder
Why it is done this way
A cut-off per target, not one for the plate, because where a Cq stops meaning “present” depends on the assay: its efficiency, its background and the purpose of the test. Choosing that value is a validation decision made per assay, often by the kit maker (Caraguel et al., 2011). Clarida applies the value you give it and records it with every saved result.
Replicates are combined by agreement rather than averaged, because near the limit of detection a target present at a few copies amplifies in some reactions and not in others, by chance alone (Forootan et al., 2017). A mean Cq would hide that; an inconclusive call shows it, and asks you to repeat the sample or decide.
A negative means little without proof that the reaction could have worked. An internal control in each sample shows that inhibition or a failed extraction did not cause it (Hoorfar et al., 2004), and no-template wells show that a positive is not contamination. MIQE 2.0 asks for both kinds of control to be reported (Bustin et al., 2025).
What it doesn’t do
| A diagnostic test | Not in Clarida: calls are for research use, such as screening panels, GMO checks, food and veterinary studies |
|---|---|
| Overruling a call by hand | Exclude a replicate, or change the target’s cut-off |
| Finding your cut-off: limit of detection, probit | Not in Clarida: you bring the cut-off, often from the kit maker |
| Calls in the Excel notebook and RDML | In the PDF and HTML report, as the call grid and counts |
Frequently asked questions
Which Ct cut-off is used for a target?
The target’s own detection cut-off first, then the experiment’s default. If neither is set, any Cq counts as detected, which is rarely what a screening run wants.
Does a failing IPC make the call invalid?
No. Clarida has no Invalid call: the sample’s row is flagged and the call stays as computed, for you to judge. A strongly positive sample that suppresses its internal control is flagged like any other failure.
Can a kit bring its own cut-offs?
Yes. A kit can bring its cut-offs into your experiment. To use another value for your run, change the target’s own cut-off: the experiment’s default never overrides it.
Are your controls called too?
No. Only unknown samples get a call, and only on targets of interest and reference genes. Controls appear as QC verdicts beside the table, and in Sample/Assay QC.