What RDES is, and the version it checks
| The format | RDES, the Real-time PCR Data Essential Spreadsheet: one tab-separated table with the raw data of one qPCR run, put forward by the HEROIC consortium as a simpler companion to RDML (Untergasser et al., 2023) |
|---|---|
| Its columns | Seven fixed ones (Well, Sample, Sample Type, Target, Target Type, Dye, then Cq or Tm); from column 8, the raw fluorescence at each cycle (each temperature for a melt); amplification and melting data in two files |
| The version checked | RDES v1.0, as in the specification at rdml.org |
| Who made the validator | Clarida, not the RDML consortium |
How it works
Drop a .tsv, .csv or .txt file of up to 10 MB; pasting the table from a spreadsheet works too. It is checked in three layers, and it is valid when all three pass.
01In the tool: RDES validator › Validation details › Format
Characters a journal’s parser can read
No quotation marks around cells, no tabs inside cells, a dot as the decimal separator, no thousand separators, and UTF-8 text. RDES needs tabs between columns, so save your table as tab-separated text before you share it.
02In the tool: RDES validator › Validation details › Structure
Every column and code as the specification spells it
- Headers: Well, Sample, Sample Type, Target, Target Type and Dye, in that order and spelled exactly, then Cq or Tm.
- The header row: cycle numbers as whole numbers, temperatures as numbers.
- Wells: letters and a number (A1, B12, AA1) with as many letters in every well; numbers alone for rotor instruments.
- Codes: sample types
unkn,ntc,nac,std,ntp,nrt,posandopt; target typestoiandref; case-sensitive. - Values: a Cq that is positive, empty, or -1.0 for a failed reaction; Tm values positive, several joined by semicolons without spaces; fluorescence as numbers.
- No empty cells in the six fixed text columns.
03In the tool: RDES validator › Validation details › Consistency
Rows that agree with each other
A sample keeps one sample type in every row, and a target keeps one target type and dye. Each combination of well, target and dye appears once, and every row has as many columns as the header.
04In the tool: RDES validator › Validation details › Download corrected RDES
Each error with its cell, and a corrected file
Every error shows where your spreadsheet would show it (Cell G2, Cq/Tm column), what is wrong and what to do; a comma decimal comes with its corrected value. A summary says what was read: the kind of data, the rows, the cycles (or temperatures) with their range, the samples, targets and dyes. When comma decimals and thousand separators are the only errors, you download a fixed tab-separated file; otherwise a partly fixed one, with a list of what is left for your source data.
What the RDES validator doesn’t do
| Judge your data | It checks the format, never your curves, Cq values or controls |
|---|---|
| Tell whether a baseline was subtracted | RDES asks for raw fluorescence; check it in your instrument’s export settings |
| Convert RDES to RDML | Not yet |
Frequently asked questions
Is the RDES validator free?
Yes, with no account: open the tool and paste your table.
Is your table stored?
No. It is sent to Clarida’s server for the check, and nothing is saved there.
Can you check a comma-separated table?
Yes, the validator reads it, but RDES itself requires tabs. Save the table as tab-separated text before you share it.
Which version of RDES does it check?
RDES v1.0, with the rules of the v1.0 specification.
Can you try it without your own data?
Yes. The Help panel has two example files to load, one with amplification data and one with melting data.